The application of various radiation detectors for brachytherapy dosimetry has motivated this study of the energy dependence of radiation quality correction factor kQ,M, the quotient of the detector responses under calibration conditions at a (60)Co unit and under the given non-reference conditions at the point of measurement, M, occurring in photon brachytherapy. The investigated detectors comprise TLD, radiochromic film, ESR, Si diode, plastic scintillator and diamond crystal detectors as well as ionization chambers of various sizes, whose measured response-energy relationships, taken from the literature, served as input data. Brachytherapy photon fields were Monte-Carlo simulated for an ideal isotropic (192)Ir point source, a model spherical (192)Ir source with steel encapsulation and a commercial HDR GammaMed Plus source.
View Article and Find Full Text PDFPurpose: The aim of this study was to investigate whether a chamber-type-specific radiation quality correction factor kQ can be determined in order to measure the reference air kerma rate of (60)Co high-dose-rate (HDR) brachytherapy sources with acceptable uncertainty by means of a well-type ionization chamber calibrated for (192)Ir HDR sources.
Methods: The calibration coefficients of 35 well-type ionization chambers of two different chamber types for radiation fields of (60)Co and (192)Ir HDR brachytherapy sources were determined experimentally. A radiation quality correction factor kQ was determined as the ratio of the calibration coefficients for (60)Co and (192)Ir.
The response of the alanine dosimeter to radiation from an Ir-192 source with respect to the absorbed dose to water, relative to Co-60 radiation, was determined experimentally as well as by Monte Carlo simulations. The experimental and Monte Carlo results for the response agree well within the limits of uncertainty. The relative response decreases with an increasing distance between the measurement volume and the source from approximately 98% at a 1 cm distance to 96% at 5 cm.
View Article and Find Full Text PDFFor medium energy x-rays produced with tube voltages from 70 to 280 kV, the absorbed dose to water, D(w), has been determined by means of water calorimetry with relative standard uncertainties ranging from 0.45% to 0.98% at 280 and 70 kV.
View Article and Find Full Text PDFRadiother Oncol
September 2011
A large air-filled parallel-plate extrapolation chamber in a phantom of water-equivalent material is used as a primary standard measuring device for low-energy interstitial brachytherapy sources from which the unit of absorbed dose to water can be derived. The chamber is suitable for low-energy photons with energies up to 50 keV. The method to determine the absorbed dose to water was newly developed based on radiation transport theory.
View Article and Find Full Text PDF