Publications by authors named "Gustavo Henrique Sumnienski Bertoldi"

Inappropriate therapy due to noise oversensing caused a true ventricular fibrillation (VF) and a life-threatening event in a patient. A 19-year-old patient with surgically corrected congenital heart disease and systolic dysfunction had an implantable cardioverter defibrillator implanted for primary prevention in 2013. This patient was admitted at the Emergency Department in June 2018 after receiving eight shocks from the device on the same day, with a prolonged syncope after the third shock.

View Article and Find Full Text PDF