In this work, 13 jet nebulizers, some of which in different configurations, were investigated in order to identify the biopharmaceutical constraints related to the quality attributes of the medicinal products, which affect their safety, efficiency, compliance, and effectiveness. The aerosolization parameters, including the aerosol output, aerosol output rate, mass median aerodynamic diameter, and fine particle fraction, were determined according to the European Standard EN 13544-1, using sodium fluoride as a reference formulation. A comparison between the aerosol output nebulization time and the fine particle fraction displayed a correlation between the aerosol quality and the nebulization rate.