This paper describes experiences in the introduction of computed radiography systems to hospitals within the West of Scotland. Setting of automatic exposure control levels has enabled dose-area product values to be maintained at 160 cGy cm² for lumbar spine antero-posterior (AP), 250 cGy cm² for lumbar spine lateral, 210 cGy cm² for pelvis AP and 260 cGy cm² for abdomen AP. The study looks at dose levels achieved after 8 y in use and compares the dose levels found from surveys undertaken in 2007 and in 2010.

View Article and Find Full Text PDF

Purpose: UVR phototherapy cabins exist in a huge variety of configurations and geometries. An accurate dose model has been developed which can be used to examine the question of cabin design to investigate factors influencing dose on a patient.

Methods: This work extends the existing dose model to entire cabins and investigates how cabin and reflector geometry influence resultant dose.

View Article and Find Full Text PDF

Purpose: Ultraviolet phototherapy is a widely used treatment which has exceptional success with a variety of skin conditions. Over-exposure to ultraviolet radiation (UVR) can however be detrimental and cause side effects such as erythema, photokeratisis, and even skin cancer. Quantifying patient dose is therefore imperative to ensure biologically effective treatment while minimizing negative repercussions.