Ultrananocrystalline diamond (UNCD), an extremely smooth, low cost diamond coating was successfully developed herein for antithrombogenic application which requires high biocompatibility, low wear, low friction, and chemical inertness. The substrate materials utilized in the Jarvik 2000 ventricular assist device (VAD), silicon carbide and titanium alloy, provide an excellent substrate match for UNCD integration. The paper addresses the development of medical-quality UNCD films to significantly improve the knowledge base regarding the defect mechanisms of UNCD films, to reduce or eliminate known wear-inducing imperfections in the film, and to thoroughly characterize and test the films as well as assembled UNCD-coated VADs.
View Article and Find Full Text PDFIn this report, an idea of integrating ultrananocrystalline diamond (UNCD) with pyrolytic carbon (PyC) -based mechanical heart valves, has been demonstrated. The report addresses the strategies to avoid graphitization and film delamination during the diamond coating. Raman and scratch tests showed that a UNCD film with high purity could adhere to the PyC substrate strongly.
View Article and Find Full Text PDF