Background: Harmonization of results allows a more effective utilization of laboratory tests; we verified the feasibility of harmonizing serum alkaline phosphatase results by two methods.
Methods: Patient sera (n=106) and candidate calibration materials (n=8) were analyzed by two methods, employing either diethanolamine (DEA) or 2-amino-2-methyl-1-propanol (AMP) as phosphate-accepting buffers. Results for patient sera by the DEA method were recalculated, with either a commutable or a non-commutable calibration material, both with values assigned by the AMP method.
Background: The availability of commutable calibrator materials may ease considerably the task of harmonizing assay results and ensuring their traceability to reference procedures. We sought to verify the commutability of potential calibrator materials and evaluate their effectiveness in harmonizing LDH results by 2 measurement methods.
Methods: We measured LDH in 109 serum samples and 31 materials, including frozen serum pools (with either normal or abnormal isoenzyme patterns), commercial stabilized materials, and the ERM-AD453/IFCC reference material.
This study was undertaken to evaluate the feasibility of using commercial control materials in a regional external quality assessment scheme (EQAS) for serum carcinoembryonic antigen (CEA) measurement. We have assessed the commutability of 12 commercial control materials using five automated immunochemical systems. We compared the intermethod behavior of the materials with that of 12-14 patient serum pools.