Combination of an ultrasonic nebulizer and plasma excitation sources for spectrochemical analysis offers desirable features of low detection limits, high sample throughput, wide dynamic range of operation, acceptable precision and accuracy, and simultaneous quantitative analytical capabilities. Moreover, the ultrasonic nebulizer does not require sample preconcentration. Recently we have developed a three-phase plasma arc (TPPA) for atomic emission spectrochemical analysis.
View Article and Find Full Text PDFFresenius J Anal Chem
December 2000
A new instrument, which uses a three-phase current to support a double-arc argon plasma torch for evaporation, atomization and excitation of solid or powder samples, is described. The sampling arc is ignited between the first and second electrode while the excitation arc is ignited between the second and third electrode. Aerosol generated from the sample (first electrode) is swept by argon gas, through a hole in the second electrode (carbon tubing electrode), into the excitation plasma.
View Article and Find Full Text PDF