Purpose: To study the impact of systematic MLC leaf positional uncertainties (stemming from mechanical inaccuracies or sub-optimal MLC modeling) on the quality of intracranial single-isocenter multi-target VMAT-SRS treatment plans. An estimation of appropriate tolerance levels is attempted.
Methods: Five patients, with three to four metastases and at least one target lying in close proximity to organs-at-risk (OARs) were included in this study.
Purpose: In the absence of a 6D couch and/or assuming considerable intrafractional patient motion, rotational errors could affect target coverage and OAR-sparing especially in multiple metastases VMAT-SRS cranial cases, which often involve the concurrent irradiation of off-axis targets. This work aims to study the dosimetric impact of rotational errors in such applications, under a comparative perspective between the single- and two-isocenter treatment techniques.
Methods: Ten patients (36 metastases) were included in this study.