J Long Term Eff Med Implants
April 2014
Using the retrieved devices from one autopsy and five reoperations, the biocompatibility of explanted Talent stent-grafts was investigated to highlight the capacity of the fabric to act as an effective scaffold to regenerate a blood conduit. The autopsy device was encapsulated both internally and externally, but the capsules did not penetrate through the fabric structure. The reoperation devices showed discrete patches of compact fibrin and irregularly scattered mural thrombi.
View Article and Find Full Text PDFJ Long Term Eff Med Implants
August 2012
Six Talent stent-grafts were harvested at reoperations (N=5) and autopsy (N=1). The explants were observed nondestructively, including gross morphology, X-rays, CT scans and closed pressure system analysis. The Nitinol frames in three devices harvested at reoperations and another harvested at autopsy were intact.
View Article and Find Full Text PDFArtif Cells Blood Substit Immobil Biotechnol
June 2009
Twenty-nine modular stent-grafts deployed transrenally to repair AAAs with short necks in dogs were harvested at autopsy of the animals after scheduled durations of implantations of 10 days, one month, three months, and six months. Analyses of the explanted devices included non-destructive techniques such as gross observations, X-rays CT scan, IVUS and angioscopy. Further to appropriate dissection, histological investigations were carried out by means of scanning electron microscopy (SEM) and light microscopy.
View Article and Find Full Text PDFNumerous endovascular stent grafts to treat intrarenal aortic aneurysms are now commercially available, and many new concepts are currently in development worldwide. In order to objectively quantify their outcomes, we propose a detailed protocol to examine a reference device that was harvested from a patient who died a few hours after endovascular stent-graft deployment for an abdominal aortic aneurysm according to the 3Bs rule (biocompatibility, biofunctionality, and biodurability). Relevant patient history of this 63-year-old man included radiotherapy treatment for lung cancer.
View Article and Find Full Text PDFArtif Cells Blood Substit Immobil Biotechnol
September 2008
Severely angulated (> 60 degrees ) or short (< 15 mm) proximal necks remain significant anatomical limitations for endovascular stent-graft repairs for abdominal aortic aneurysms. Ensuring proper proximal fixation of the stent-graft to the host artery without the short-or long-term risks of endoleak or migration represents a particular technical challenge for these anatomical circumstances. An innovative balloon expandable stent combined with a weft-knitted prosthesis was specifically designed for these situations by modelling the stent to the neck anatomy without overdistension or potential barotrauma allowing better incorporation of the device.
View Article and Find Full Text PDFThe arterial wall viscoelasticity plays an essential role in the vascular responsiveness to vasoactive drugs or pathologies. The aim of this investigation was to derive and compare resonance curve (RC), natural frequency (f(0)), dynamic modulus of elasticity (E'), and coefficient of viscosity (beta) of (i) vital and devitalized preparations of rat thoracic and abdominal aorta, (ii) human arterial prostheses, and to study the histomorphology of vital and devitalized rat aorta. The method of low frequency forced oscillations was employed.
View Article and Find Full Text PDFA patient was fitted with an Anaconda stent graft for which there was a persistent type II endoleak. Two subsequent attempts at embolization were unable to resolve the endoleak. The diameter of the aneurysm varied initially from 5.
View Article and Find Full Text PDFFurther to the rapid enlargement of an aneurysm to 5.6 cm in diameter after 3 years of surveillance, a 79-year-old patient was fitted with a Vanguard modular stent graft and monitored on a regular basis for 6 years. Two years later, the aneurysmal sac ruptured.
View Article and Find Full Text PDFPurpose: To validate the deployment, in vivo performance, biostability, and healing capacity of the Anaconda self-expanding endoprosthesis in a canine aortic aneurysm model.
Methods: Aneurysms were surgically created in 12 dogs by sewing a woven polyester patch onto the anterior side of the thoracic or abdominal aorta. Anaconda prostheses were implanted transfemorally for prescheduled periods (1 or 3 months).