A 48-year-old man was admitted after an episode of aborted sudden death with external defibrillation. An implantable cardioverter defibrillator implanted 2 years before for secondary prevention failed to sense properly an episode of ventricular fibrillation. Interrogation of the device showed large oscillatory changes of the amplitude of the local electrogram during ventricular fibrillation, causing undersensing and inappropriate refraining from shock therapy.