This study aims to optimize a single preparation methodology based on solid-phase extraction (SPE) that could fit both target and non-target screening of organic biomarkers in raw wastewater, allowing the cross-comparison of results obtained from a same dataset. The efficiency of SPE sorbents used alone (HLB) or in combination in a multilayer cartridge was evaluated based on (i) the extraction recovery and matrix effect in environmental samples (surface water and wastewater) for a list of biomarkers (pharmaceuticals, licit and illicit drugs, artificial sweeteners, isoprostanes, polyphenols) and (ii) a number of detected features and their intensity in HRMS. The selected method uses a combination of three SPE sorbents mixed together (HLB, X-AW and X-CW) and seems to take full advantage of each, providing satisfactory validation parameters (recovery, instrumental limit of detection, linearity range and limit of quantification) over a large range of physico-chemical properties while ensuring promising results for non-target screening applications.
View Article and Find Full Text PDF