Pacing Clin Electrophysiol
November 1996
Using telemetry, right atrial electrogram (RA), and marker channel of atrial sense events (MA) in combination with the left atrial electrogram (LA), recorded by a filtered bipolar esophageal lead, interatrial conduction during submaximal exercise and at rest was examined in 46 DDD pacemaker patients. The RA-LA and MA-LA conduction times measured in the presence of atrial sensing (VDD) as well as the conduction time SA-LA from atrial stimulus (SA) to LA, determined during atrial pacing (DDD) were found to be individual constants independent of exercise induced sympathetic influences. Thus, having determined an optimal mechanical interval (LA-LV)mech/opt from left atrium to ventricle by other methods, the optimal AV delay for DDD as well as for VDD operation can be calculated by the sum of the appropriate interatrial conduction time (SA-LA, respectively MA-LA) and the (LA-LV)mech/opt interval.
View Article and Find Full Text PDF