Purpose: To evaluate the impact of tuning the beam configurations parameters on the Analytical Anisotropic Algorithm (AAA) and the Acuros XB (AXB) algorithm for small fields using Monte Carlo simulations and measurements.
Methods: The TrueBeam STx with the high-definition 120 multi-leaf collimator (HD120-MLC) was modeled with Geant4 application for emission tomography (GATE) Monte Carlo platform and validated against measurements. The impact of varying the effective spot size (ESS) and dosimetric leaf gap (DLG) on AAA and AXB calculations was carried out for small MLC-fields ranging from 0.