Chemical shift imaging benefits from signal-to-noise ratio (SNR) and chemical shift dispersion increases at stronger main field such as 7 Tesla, but the associated shorter radiofrequency (RF) wavelengths encountered require B1+ mitigation over both the spatial field of view (FOV) and a specified spectral bandwidth. The bandwidth constraint presents a challenge for previously proposed spatially tailored B1+ mitigation methods, which are based on a type of echovolumnar trajectory referred to as "spokes" or "fast-kz". Although such pulses, in conjunction with parallel excitation methodology, can efficiently mitigate large B1+ inhomogeneities and achieve relatively short pulse durations with slice-selective excitations, they exhibit a narrow-band off-resonance response and may not be suitable for applications that require B1+ mitigation over a large spectral bandwidth.
View Article and Find Full Text PDFJ Magn Reson Imaging
January 2009
Purpose: To clarify whether sinusoidal pulses possess lower thresholds than rectangular ones at perception threshold, a statement often made that contradicts the theory of stimulation.
Materials And Methods: The results of a nerve stimulation study with 65 volunteers and with trapezoidal and sinusoidal gradient pulses were used to apply the combination of the electric field, induced in the tissue of the human body, with the "Fundamental Law of Electrostimulation." This law claims that the waveshape of a pulse is not essential as long as the amplitude of the pulse does not decrease below rheobase (rheobase condition).
Slice-selective RF waveforms that mitigate severe B1+ inhomogeneity at 7 Tesla using parallel excitation were designed and validated in a water phantom and human studies on six subjects using a 16-element degenerate stripline array coil driven with a butler matrix to utilize the eight most favorable birdcage modes. The parallel RF waveform design applied magnitude least-squares (MLS) criteria with an optimized k-space excitation trajectory to significantly improve profile uniformity compared to conventional least-squares (LS) designs. Parallel excitation RF pulses designed to excite a uniform in-plane flip angle (FA) with slice selection in the z-direction were demonstrated and compared with conventional sinc-pulse excitation and RF shimming.
View Article and Find Full Text PDFAt high magnetic field, B(1)(+) non-uniformity causes undesired inhomogeneity in SNR and image contrast. Parallel RF transmission using tailored 3D k-space trajectory design has been shown to correct for this problem and produce highly uniform in-plane magnetization with good slice selection profile within a relatively short excitation duration. However, at large flip angles the excitation k-space based design method fails.
View Article and Find Full Text PDFAn eight-rung, 3T degenerate birdcage coil (DBC) was constructed and evaluated for accelerated parallel excitation of the head with eight independent excitation channels. Two mode configurations were tested. In the first, each of the eight loops formed by the birdcage was individually excited, producing an excitation pattern similar to a loop coil array.
View Article and Find Full Text PDFSpatially selective RF waveforms were designed and demonstrated for parallel excitation with a dedicated eight-coil transmit array on a modified 3T human MRI scanner. Measured excitation profiles of individual coils in the array were used in a low-flip-angle pulse design to achieve desired spatial target profiles with two- (2D) and three-dimensional (3D) k-space excitation with simultaneous transmission of RF on eight channels. The 2D pulse excited a high-resolution spatial pattern in-plane, while the 3D trajectory produced high-quality slice selection with a uniform in-plane excitation despite the highly nonuniform individual spatial profiles of the coil array.
View Article and Find Full Text PDFPurpose: To increase gradient stimulation from 100% to a fixed 120% level and to assess patient acceptance of the degree of peripheral nerve stimulation (PNS) at magnetic resonance (MR) imaging.
Materials And Methods: Two hundred ten patients underwent MR imaging of various body regions according to clinical indications. An additional three-dimensional fast low-angle shot sequence with the 120% stimulation level was performed.
Purpose: To compare thresholds for peripheral nerve stimulation from gradient switching in whole body magnetic resonance (MR) equipment of different design.
Materials And Methods: Threshold data obtained in three experiments were reformatted into a single joint data set describing thresholds for anterio-posterior (AP) gradient orientation and Echo Planar Imaging (EPI) waveforms with bipolar ramp times between 0.07 and 1.