This work presents an algorithm for the detection and classification of QRS complexes based on the continuous wavelet transform (CWT) with splines. This approach can evaluate the CWT at any integer scale and the analysis is not restricted to powers of two. The QRS detector comprises four stages: implementation of CWT with splines, detection of QRS complexes, searching for undetected QRS complexes, and correction of the R wave peak location in detected QRS complexes.

View Article and Find Full Text PDF

A low-power long-term ambulatory ECG monitor was developed for the acquisition, storage and processing of three simultaneous leads DI, aVF and V2 with a beat-to-beat heart rate measurement in real time. It provides long-term continuous ECG recordings until 84 h. The monitor uses a QRS complex detection algorithm based on the continuous wavelet transform with splines, which automatically selects the scale for the analysis of ECG records with different sampling frequencies.

View Article and Find Full Text PDF

Cardiovascular diseases are currently the leading cause of death worldwide. Thus, there is a need for non-invasive ambulatory (Holter) ECG monitors with automatic measurements of ECG intervals to evaluate electrocardiographic abnormalities of patients with cardiac diseases. This work presents the implementation of algorithms in an FPGA for beat-to-beat heart rate and RT interval measurements based on the continuous wavelet transform (CWT) with splines for a prototype of an ambulatory ECG monitor of three leads.

View Article and Find Full Text PDF