Introduction: Current gold standard for the treatment of symphyseal disruptions includes anterior plating, almost entirely prohibiting symphyseal mobility and resulting in an iatrogenic arthrodesis followed by high rates of implant failure. Minimally invasive tape suture constructs have been found to maintain the micro mobility of ligamentous injuries, yet still providing sufficient biomechanical stability. Recently, this technique has been primarily investigated for symphyseal disruptions on synthetic pelvic models.
View Article and Find Full Text PDFPurpose: Open reduction and internal fixation using anterior plate osteosynthesis currently represents the gold standard for the treatment of symphyseal disruptions. Since postoperative screw loosening with consequent implant failure is frequently observed, this study aims to evaluate if and to what extent augmentation can increase the pull-out force of symphyseal screws to improve the constructs stability.
Methods: Twelve human cadaveric anterior pelvic rings were separated at the symphyseal joint for bilateral testing, consequently achieving comparable sites.
Purpose: Osteoporotic bone tissue appears to be an important risk factor for implant loosening, compromising the stability of surgical implants. However, it is unclear whether lumbar measured bone mineral density (BMD) is of any predictive value for stability of surgical implants at the pubic symphysis. This study examines the fixation strength of cortical screws in human cadaver specimens with different BMDs.
View Article and Find Full Text PDFIntroduction: Open book fractures are challenging injuries oftentimes requiring surgical treatment. The current treatment of choice is symphyseal plating, which requires extensive surgery and entirely limits physiological movement of the symphyseal joint, frequently resulting in implant failure. Therefore, we investigated the biomechanical properties of a semi-rigid implant (modified SpeedBridge) as a minimally invasive tape suture construct for the treatment of open book fractures and evaluated the superiority of two techniques of implementation: criss-cross vs.
View Article and Find Full Text PDF