In this study, we discuss the origin of the slightly increased response of the charged aerosol detector when low-concentration polar drugs formulated with sodium chloride are analyzed by hydrophilic interaction liquid chromatography coupled to the charged aerosol detector. In the case of tromethamine mixed with saline solutions, we investigated several levels including the mobile phase, sample matrix, and detection. We show that the analysis of the rich-salted sample results in both interactions with the mobile phase modifiers and the stationary phase during the run time.
View Article and Find Full Text PDFCodeine N-oxide 2 is an active metabolite of codeine obtained by oxidation and observed as a degradant in codeine drug products such as syrups. Oxidation of codeine's N-methyl function can deliver two regio-isomers due to chirality of the tetra-substituted nitrogen. Hydrogen peroxide oxidation of codeine was performed and induced two different isomers in a 9:1 ratio; these isomers were isolated using preparative high performance liquid chromatography (HPLC) and fully characterized by nuclear magnetic resonance (NMR) techniques.
View Article and Find Full Text PDFTromethamine (TMM), often encountered in a final drug product, exhibits interesting chemical properties as a counter ion, buffer, or active ingredient. European and US pharmacopeias propose titration against hydrogen chloride for TMM assays. However, this method can be a hindrance when using drugs containing low concentrations of TMM in complex buffered formulations.
View Article and Find Full Text PDF