The effects of extra-column band spreading, LC system operating pressure, and separation temperature were investigated for sub-2-microm particle columns using both a conventional HPLC system as well as a UPLC system. The contributions of both volume- and time-based extra-column effects were analyzed in detail. In addition, the performance difference between columns containing 2.
View Article and Find Full Text PDFAn un-derivatized 1.7 microm ethylene bridged hybrid (BEH) particle was evaluated for its utility in retaining polar species in hydrophilic interaction chromatography (HILIC), and was compared to a 3 microm un-derivatized silica material. Retentivity as a function of mobile phase pH, polar modifier and ACN content was examined.
View Article and Find Full Text PDFThe effect of combining sub-2 microm porous particles with elevated operating temperatures on chromatographic performance has been investigated in terms of chromatographic efficiency, productivity, peak elution order, and observed operating pressure. The use of elevated temperature in LC does not increase the obtainable performance but allows the same performance to be obtained in less time. Increasing the column temperature did allow the use of longer columns, generating column efficiencies in excess of 100,000 plates and gradient peak capacities approaching 1000.
View Article and Find Full Text PDF