The response range of an ion-selective electrode (ISE) has been described by counterion interference at the lower and Donnan failure at the upper detection limit. This approach fails when the potentiometric response at the upper detection limit exhibits an apparently super-Nernstian response, as has been reported repeatedly for H-selective electrodes. While also observed when samples contain other anions, super-Nernstian responses at low pH are a problem in particular for samples that contain phthalate, a common component of commercial pH calibration solutions.
View Article and Find Full Text PDFAgCl-coated silver fabricated with the thermal-electrolytic method can be used to prepare more reproducible reference electrodes than Ag/AgCl prepared with alternative methods such as electrolytic and chemical AgCl deposition or thermal fabrication. However, thermal-electrolytic fabrication requires a scaffold material upon which to build the layers upon. Platinum and rhodium have been used for this purpose as they are mechanically strong and chemically inert, but their cost is prohibitive for wider application.
View Article and Find Full Text PDF