Aims: Videodensitometric assessment of aortography provides a periprocedural quantitation of prosthetic valve regurgitation (PVR) after transcatheter aortic valve implantation. We sought to compare the videodensitometric parameters of PVR severity to the regurgitation fraction (RF) in a controlled in vitro setting.
Methods And Results: In a mock circulation system, a transcatheter balloon-expandable valve inserted at the aortic valve position was gradually deformed to induce different grades of paravalvular leakage and the RF was measured with a transonic flow probe.
Aims: In the minimalist transcatheter aortic valve implantation (TAVI) era, the usage of transoesophageal echocardiography has become restricted. Conversely, aortography has gained clinical ground in quantifying prosthetic valve regurgitation (PVR) during the procedure. In a mock circulation system, we sought to compare the contrast volume required and the accuracy of aortographic videodensitometric PVR assessment using a synchronised diastolic and standard (non-synchronised) injection aortography.