The desire for portable Raman spectrometers is continuously driving the development of novel spectrometer architectures where miniaturisation can be achieved without the penalty of a poorer detection performance. Spatial heterodyne spectrometers are emerging as potential candidates for challenging the dominance of traditional grating spectrometers, thanks to their larger etendue and greater potential for miniaturisation. This paper provides a generic analytical model for estimating and comparing the detection performance of Raman spectrometers based on grating spectrometer and spatial heterodyne spectrometer designs by deriving the analytical expressions for the performance estimator (signal-to-noise ratio, SNR) for both types of spectrometers.

View Article and Find Full Text PDF

A high power continuous wave quantum cascade laser operating around 1900 cm(-1) has been used to conduct Lamb dip spectroscopy on a low pressure sample of NO. The widths of the Lamb dips indicate that the laser linewidth is 800 ± 60 kHz and the power sufficient to induce significant population transfer of up to 35%. While the Lamb dip signals are symmetric at low laser chirp rates, they become increasingly asymmetric as the chirp rate increases, further confirming the significant degree of population transfer.