The output signal of a Kovasznay type vorticity probe is in first-order approximation proportional to the longitudinal component of the vorticity omegax= partial differentialw/ partial differentialy- partial differentialv/ partial differentialz and does not depend on the two transverse components of the flow velocity v, w. An experimental investigation of the influence of all three fluctuating velocity components on the longitudinal vorticity signal showed that their influence may not be neglected. The error in the probe response caused by the longitudinal component of the flow velocity was easily corrected using the instantaneous longitudinal velocity component and digital measuring techniques.
View Article and Find Full Text PDF