Objective: Recent studies have renewed interest in reflectance pulse oximetry, specifically for monitoring the patient's forehead. Blood circulation on the forehead immediately above the eyebrow is fed by arteries that branch from the internal carotid artery and lack the vasoconstrictor response present in more peripheral regions. Some investigators question, however, the reliability of monitoring SpO2 on the forehead due to prior reported inaccurate readings with reflectance sensors.

View Article and Find Full Text PDF

This study aimed to investigate and compare the performance of the algorithms contained in the newest generation of pulse oximeters (Masimo SET in IVY2000, Nellcor Oxismart N-3000, Agilent M3 rev. B) against a traditional pulse oximeter (Agilent CMS rel. A.

View Article and Find Full Text PDF

Objective: To develop and test a method for standardized calibration of pulse oximeters.

Methods: A novel pulse oximeter calibration technique capable of simulating the behavior of real patients is discussed. It is based on an artificial finger with a variable spectral-resolved light attenuator in conjunction with an extensive clinical database of time-resolved optical transmission spectra of patients fingers in the wavelength range 600-1000 nm.

View Article and Find Full Text PDF

Objective: There is no commonly accepted in vivo calibration method for pulse oximeters available up to now. On the basis of a prototype device for the calibration of pulse oximeters which was introduced recently, a second approach based on the same concept was tackled in order to design a reliable method for standardized calibration of pulse oximeters.

Methods: An extensive clinical database of time-resolved optical transmission spectra of patient fingers is used to simulate the behavior of patients.

View Article and Find Full Text PDF