Most of the treatment units, both new and old models, are equipped with a megavoltage portal imager but its use for volumetric imaging is limited. This is mainly due to the poor image quality produced by the high-energy treatment beam (> 6MV). A linac at our center is equipped with a prototype 2.
View Article and Find Full Text PDFPurpose: To evaluate the use of megavoltage cone-beam computed tomography (MV CBCT) to measure interfractional variation in lung tumor position.
Methods And Materials: Eight non-small-cell lung cancer patients participated in the study, 4 with respiratory gating and 4 without. All patients underwent MV CBCT scanning at weekly intervals.
We report on the capabilities of a low-dose megavoltage cone-beam computed tomography (MV CBCT) system. The high-efficiency image receptor consists of a photodiode array coupled to a scintillator composed of individual CsI crystals. The CBCT system uses the 6 MV beam from a linear accelerator.
View Article and Find Full Text PDFWe have previously described a low-dose megavoltage cone beam computed tomography (MV CBCT) system capable of producing projection image using one beam pulse. In this study, we report on its integration with respiratory gating for gated radiotherapy. The respiratory gating system tracks a reflective marker on the patient's abdomen midway between the xiphoid and umbilicus, and disables radiation delivery when the marker position is outside predefined thresholds.
View Article and Find Full Text PDFBackground And Purpose: To implement an on-line correction scheme based on implanted markers to reduce treatment margins in external beam radiation therapy (EBRT) of carcinoma of the prostate. In turn reduction in treatment margins reduces irradiated volumes and offers the possibility of reduced normal tissue complications or escalated target dose.
Patients And Methods: Five or six gold markers were implanted in 10 patients treated for prostate carcinoma using EBRT.