Aims: Pulsed-field ablation (PFA) is an emerging non-thermal ablation method based on the biophysical phenomenon of electroporation. Data on PFA cardiac selectivity nature and tissue-specific thresholds are lacking. We aim to compare the in vivo differential effect of high-frequency irreversible electroporation (HF-IRE) protocols on various tissues.
View Article and Find Full Text PDFThis study compared monophasic 100-μs pulses with high-frequency electroporation (HF-EP) bursts using an in vivo animal model. Myocardial damage was evaluated by histologic analysis. Compared with 10 monophasic pulses, 20 bursts of HF-EP at 100 and 150 kHz were associated with less damage.
View Article and Find Full Text PDF