We evaluated a new protocol for measurement of cyclosporine A (CsA) 2 H after dose (C2) on the V-Twin® analyzer. Imprecision, recovery, and linearity were determined using CsA-spiked blood pools. Accuracy was evaluated using specimens from renal, cardiac, and liver transplant patients, and results were compared with those from liquid chromatography-tandem mass spectrometry (LC-MS/MS) and the Abbott TDx®/TDxFLx® assay.
View Article and Find Full Text PDFObjectives: This study evaluated the quantitative total betahCG assay on the Stratus CS point-of-care instrument at two medical centers.
Design And Methods: Analytical sensitivity, linearity, within-run and total imprecision, interferences, dilution recovery, method comparison (Dimension RxL), comparison of matched heparinized whole blood and plasma samples, and determination of the normal reference interval were studied.
Results: Analytical sensitivity was <0.