This article aims to present the development, construction, and evaluation of an alternative computed tomography dose index (CTDI) phantom. Epoxy resin was mixed with an iodine-based contrast agent to produce radiological characteristics resembling polymethyl methacrylate (PMMA) as a standard CTDI phantom. As a preliminary study, testing was carried out using computed tomography images (80 and 120 kVp) on 12 variations of epoxy-iodine resin mixtures to obtain relative electron density () values and effective atomic numbers () of the samples.
View Article and Find Full Text PDFPurpose: This study aims to assess the accuracy of a modified electron beam calibration based on the IAEA TRS-398 and AAPM-TG-51 in multicenter radiotherapy.
Methods: This study was performed using the Elekta and Varian Linear Accelerator electron beams with energies of 4-22 MeV under reference conditions using cylindrical (PTW 30013, IBA FC65-G, and IBA FC65-P) and parallel-plate (PTW 34045, PTW 34001, and IBA PPC-40) chambers. The modified calibration used a cylindrical chamber and an updated based on Monte Carlo calculations, whereas TRS-398 and TG-51 used cylindrical and parallel-plate chambers for reference dosimetry.