Purpose: To develop a novel, monthly quality assurance (QA) regimen for a proton therapy system that uses 2 custom phantoms, each housing a commercial scintillator detector and a charge-coupled device camera. The novel metrology system assessed QA trends at a pediatric proton therapy center from 2018 to 2022.

Materials And Methods: The measurement system was designed to accommodate horizontal and vertical positioning of the commercial device and to enable gantry and couch isocentricity measurements (using a star shot procedure), proton spot profile verification, and imaging and radiation congruence tests to be performed simultaneously in the dual-phantom setup.

View Article and Find Full Text PDF

Purpose/objective(s): Surface-guided radiation therapy (SGRT) can track the patient surface noninvasively to complement radiographic image-guided radiation therapy with a standard 3-camera system and a single radiation/image isocenter. Here we report the commissioning of a novel SGRT system that monitors three imaging isocenters locations in a proton half-gantry room with a unique 5-camera configuration.

Materials/methods: The proton half-gantry room has three image isocenters, designated ISO-0, ISO-1, and ISO-2, to cover various anatomical sites via a robotic ceiling-mounted cone-beam CT.

View Article and Find Full Text PDF

We introduce a custom-made silicone-filled vaginal spacer for use during treatment of female patients receiving pelvic proton radiation therapy. Commercially available vaginal dilators can be purchased as hollow objects; when filled with a media, they can act as a beam stopper and/or tissue spacer while pushing uninvolved vaginal wall away from a high-dose region. Dosimetric advantages of these specifically constructed silicone-filled vaginal spacers were investigated when compared to the unaltered commercially available product or no vaginal spacer in pediatric proton therapy.