Currently, there are multiple breast dosimetry estimation methods for mammography and its variants in use throughout the world. This fact alone introduces uncertainty, since it is often impossible to distinguish which model is internally used by a specific imaging system. In addition, all current models are hampered by various limitations, in terms of overly simplified models of the breast and its composition, as well as simplistic models of the imaging system.
View Article and Find Full Text PDFWe set out a fully developed algorithm for adapting mammography images to appear as if acquired using different technique factors by changing the signal and noise within the images. The algorithm accounts for difference between the absorption by the object being imaged and the imaging system. Images were acquired using a Hologic Selenia Dimensions x-ray unit for the validation, of three thicknesses of polymethyl methacrylate (PMMA) blocks with or without different thicknesses of PMMA contrast objects acquired for a range of technique factors.
View Article and Find Full Text PDFBackground: Understanding the magnitude and variability of the radiation dose absorbed by the breast fibroglandular tissue during mammography and digital breast tomosynthesis (DBT) is of paramount importance to assess risks versus benefits. Although homogeneous breast models have been proposed and used for decades for this purpose, they do not accurately reflect the actual heterogeneous distribution of the fibroglandular tissue in the breast, leading to biases in the estimation of dose from these modalities.
Purpose: To develop and validate a method to generate patient-derived, heterogeneous digital breast phantoms for breast dosimetry in mammography and DBT.
Purpose: The purpose of this study was to measure the threshold diameter of calcifications and masses for 2D imaging, digital breast tomosynthesis (DBT), and synthetic 2D images, for a range of breast glandularities. This study shows the limits of detection for each of the technologies and the strengths and weaknesses of each in terms of visualizing the radiological features of small cancers.
Methods: Mathematical voxel breast phantoms with glandularities by volume of 9%, 18%, and 30% with a thickness of 53 mm were created.
Objectives: This study was designed to compare the detection of subtle lesions (calcification clusters or masses) when using the combination of digital breast tomosynthesis (DBT) and synthetic mammography (SM) with digital mammography (DM) alone or combined with DBT.
Methods: A set of 166 cases without cancer was acquired on a DBT mammography system. Realistic subtle calcification clusters and masses in the DM images and DBT planes were digitally inserted into 104 of the acquired cases.
Purpose: To develop a patient-based breast density model by characterizing the fibroglandular tissue distribution in patient breasts during compression for mammography and digital breast tomosynthesis (DBT) imaging.
Methods: In this prospective study, 88 breast images were acquired using a dedicated breast computed tomography (CT) system. The breasts in the images were classified into their three main tissue components and mechanically compressed to mimic the positioning for mammographic acquisition of the craniocaudal (CC) and mediolateral oblique (MLO) views.
Purpose: Virtual clinical trials (VCT) are a powerful imaging tool that can be used to investigate digital breast tomosynthesis (DBT) technology. In this work, a fast and simple method is proposed to estimate the two-dimensional distribution of scattered radiation which is needed when simulating DBT geometries in VCTs.
Methods: Monte Carlo simulations are used to precalculate scatter-to-primary ratio (SPR) for a range of low-resolution homogeneous phantoms.
Digital breast tomosynthesis (DBT) is currently under consideration for replacement of, or combined use with 2D-mammography in national breast screening programmes. To investigate the potential benefits that DBT can bring to screening, the threshold detectable lesion diameters were measured for different forms of DBT in comparison to 2D-mammography. The aim of this study was to compare the threshold detectable mass diameters obtained with narrow angle (15°/15 projections) and wide angle (50°/25 projections) DBT in comparison to 2D-mammography.
View Article and Find Full Text PDFKnowledge of x-ray attenuation is essential for developing and evaluating x-ray imaging technologies. In mammography, measurement of breast density, dose estimation, and differentiation between cysts and solid tumours are example applications requiring accurate data on tissue attenuation. Published attenuation data are, however, sparse and cover a relatively wide range.
View Article and Find Full Text PDFThis work investigates the detection performance of specialist and non-specialist observers for different targets in 2D-mammography and digital breast tomosynthesis (DBT) using the OPTIMAM virtual clinical trials (VCT) Toolbox and a 4-alternative forced choice (4AFC) assessment paradigm. Using 2D-mammography and DBT images of virtual breast phantoms, we compare the detection limits of simple uniform spherical targets and irregular solid masses. Target diameters of 4 mm and 6 mm have been chosen to represent target sizes close to the minimum detectable size found in breast screening, across a range of controlled contrast levels.
View Article and Find Full Text PDFPurpose: Model observers (MOs) are of interest in the field of medical imaging to assess image quality. However, before procedures using MOs can be proposed in quality control guidelines for mammography systems, we need to know whether MOs are sensitive to changes in image quality and correlations in background structure. Therefore, as a proof of principle, in this study human and model observer (MO) performance are compared for the detection of calcification-like objects using different background structures and image quality levels of unprocessed mammography images.
View Article and Find Full Text PDFPurpose: To demonstrate a method of simulating mammography images of the CDMAM phantom and to investigate the coefficient of variation (CoV) in the threshold gold thickness (t) measurements associated with use of the phantom.
Methods: The noise and sharpness of Hologic Dimensions and GE Essential mammography systems were characterized to provide data for the simulation. The simulation method was validated by comparing the t results of real and simulated images of the CDMAM phantom for three different doses and the two systems.
Purpose: To characterize the dependence of normalized glandular dose (DgN) on various breast model and image acquisition parameters during spot compression mammography and other partial breast irradiation conditions, and evaluate alternative previously proposed dose-related metrics for this breast imaging modality.
Methods: Using Monte Carlo simulations with both simple homogeneous breast models and patient-specific breasts, three different dose-related metrics for spot compression mammography were compared: the standard DgN, the normalized glandular dose to only the directly irradiated portion of the breast (DgNv), and the DgN obtained by the product of the DgN for full field irradiation and the ratio of the mid-height area of the irradiated breast to the entire breast area (DgN ). How these metrics vary with field-of-view size, spot area thickness, x-ray energy, spot area and position, breast shape and size, and system geometry was characterized for the simple breast model and a comparison of the simple model results to those with patient-specific breasts was also performed.
A novel method has been developed for generating quasi-realistic voxel phantoms which simulate the compressed breast in mammography and digital breast tomosynthesis (DBT). The models are suitable for use in virtual clinical trials requiring realistic anatomy which use the multiple alternative forced choice (AFC) paradigm and patches from the complete breast image. The breast models are produced by extracting features of breast tissue components from DBT clinical images including skin, adipose and fibro-glandular tissue, blood vessels and Cooper's ligaments.
View Article and Find Full Text PDFIn-depth evaluation of the noise and sharpness characteristics of FujiFilm Innovality, GE SenoClaire, Hologic Selenia Dimensions and Siemens Inspiration digital breast tomosynthesis (DBT) systems was performed with the intention of improving image simulation for virtual clinical trials. Noise power spectra (NPS) and modulation transfer function curves (MTF) were measured for planar modes and for the first and central projections for DBT modes. In DBT mode, the x-ray beam was blocked for the projections before the central projection in order to remove the influence of lag and ghosting from the previous images.
View Article and Find Full Text PDFDigital breast tomosynthesis (DBT) is under consideration to replace or to be used in combination with 2D-mammography in breast screening. The aim of this study was the comparison of the detection of microcalcification clusters by human observers in simulated breast images using 2D-mammography, narrow angle (15°/15 projections) and wide angle (50°/25 projections) DBT. The effects of the cluster height in the breast and the dose to the breast on calcification detection were also tested.
View Article and Find Full Text PDFObjective: To estimate the risks and benefits of breast screening in terms of number of deaths due to radiation-induced cancers and the number of lives saved owing to modern screening in the National Health Service Breast Screening Programme (NHSBSP) in England.
Methods: Radiation risk model, patient dose data and data from national screening statistics were used to estimate the number of deaths due to radiation-induced breast cancers in the NHSBSP in England. Dose and dose effectiveness factors (DDREFs) equal to one and two were assumed.
The estimation of the mean glandular dose to the breast (MGD) for x-ray based imaging modalities forms an essential part of quality control and is needed for risk estimation and for system design and optimisation. This review considers the development of methods for estimating the MGD for mammography, digital breast tomosynthesis (DBT) and dedicated breast CT (DBCT). Almost all of the methodology used employs Monte Carlo calculated conversion factors to relate the measurable quantity, generally the incident air kerma, to the MGD.
View Article and Find Full Text PDFPurpose: To investigate the relationship between image quality measurements and the clinical performance of digital mammographic systems.
Methods: Mammograms containing subtle malignant non-calcification lesions and simulated malignant calcification clusters were adapted to appear as if acquired by four types of detector. Observers searched for suspicious lesions and gave these a malignancy score.
Knowledge of x-ray attenuation is essential for developing and evaluating x-ray imaging technologies. For instance, techniques to distinguish between cysts and solid tumours at mammography screening would be highly desirable to reduce recalls, but the development requires knowledge of the x-ray attenuation for cysts and tumours. We have previously measured the attenuation of cyst fluid using photon-counting spectral mammography.
View Article and Find Full Text PDFObjective: To compare the performance of different types of detectors in breast cancer detection.
Methods: A mammography image set containing subtle malignant non-calcification lesions, biopsy-proven benign lesions, simulated malignant calcification clusters and normals was acquired using amorphous-selenium (a-Se) detectors. The images were adapted to simulate four types of detectors at the same radiation dose: digital radiography (DR) detectors with a-Se and caesium iodide (CsI) convertors, and computed radiography (CR) detectors with a powder phosphor (PIP) and a needle phosphor (NIP).
This paper introduces a hybrid method for performing detection studies in projection image based modalities, based on image acquisitions of target objects and patients. The method was used to compare 2D mammography and digital breast tomosynthesis (DBT) in terms of the detection performance of spherical densities and microcalcifications. The method starts with the acquisition of spheres of different glandular equivalent densities and microcalcifications of different sizes immersed in a homogeneous breast tissue simulating medium.
View Article and Find Full Text PDFPurpose: The aim of this work is to create a model to predict the noise power spectra (NPS) for a range of mammographic radiographic factors. The noise model was necessary to degrade images acquired on one system to match the image quality of different systems for a range of beam qualities.
Methods: Five detectors and x-ray systems [Hologic Selenia (ASEh), Carestream computed radiography CR900 (CRc), GE Essential (CSI), Carestream NIP (NIPc), and Siemens Inspiration (ASEs)] were characterized for this study.
Purpose: This work proposes a new method of building 3D breast mass models with different morphological shapes and describes the validation of the realism of their appearance after simulation into 2D digital mammograms and breast tomosynthesis images.
Methods: Twenty-five contrast enhanced MRI breast lesions were collected and each mass was manually segmented in the three orthogonal views: sagittal, coronal, and transversal. The segmented models were combined, resampled to have isotropic voxel sizes, triangularly meshed, and scaled to different sizes.
OBJECTIVE. The objective of our study was to investigate the effect of image processing on the detection of cancers in digital mammography images. MATERIALS AND METHODS.
View Article and Find Full Text PDF