Background: Longer term outcomes of transcatheter aortic valve replacement (TAVR) with the SAPIEN 3 Ultra RESILIA (S3UR) valve over its predecessors have not been reported in a national registry.
Objectives: The aim of this study was to compare the 1-year clinical and echocardiographic outcomes of the S3UR with those of the SAPIEN 3 (S3) and SAPIEN 3 Ultra (S3U) in the Society of Thoracic Surgeons/American College of Cardiology TVT (Transcatheter Valve Therapy) Registry.
Methods: From September 2022 to March 2023, patients who underwent native TAVR with the S3UR or the S3 or S3U valve were propensity matched and compared.
Background: The fifth-generation SAPIEN 3 Ultra Resilia valve (S3UR) incorporates several design changes as compared with its predecessors, the SAPIEN 3 (S3) and SAPIEN 3 Ultra (S3U) valves, including bovine leaflets treated with a novel process intended to reduce structural valve deterioration via calcification, as well as a taller external skirt on the 29-mm valve size to reduce paravalvular leak (PVL). The clinical performance of S3UR compared with S3 and S3U in a large patient population has not been previously reported.
Objectives: The aim of this study was to compare S3UR to S3/S3U for procedural, in-hospital, and 30-day clinical and echocardiographic outcomes after transcatheter aortic valve replacement (TAVR).