Publications by authors named "David J Langton"

Total Knee Replacements (TKRs) are a commonly used treatment to help patients suffering from severely damaged knee joints, which is normally brought on by osteoarthritis. The aim of the surgery is to reduce pain and regain function of the joint, however, some of these implants fail prematurely with implant wear being one of the main factors of failure. Computational analysis is an efficient tool that can provide an in-depth insight on the evolution of wear, before utilising experimental techniques which are time-consuming and costly.

View Article and Find Full Text PDF

Total Hip Arthroplasty has been a revolutionary technique in restoring mobility to patients with damaged hip joints. The introduction of modular components of the hip prosthesis allowed for bespoke solutions based on the requirements of the patient. The femoral stem is designed with a conical trunnion to allow for assembly of different femoral head sizes based on surgical requirements.

View Article and Find Full Text PDF

Background: We investigated the relationship between the backside deformation of polyethylene (PE) tibial inserts and aseptic loosening of the Option stemmed tibial tray used with Zimmer NexGen posterior-stabilised (PS) devices. We hypothesized that explanted inserts used in PS designs would exhibit greater extents of PE backside deformation than those used in equivalent cruciate retaining (CR) designs and that PE inserts retrieved from total knee arthroplasties (TKAs) revised for aseptic tibial tray loosening would exhibit greater extents of backside deformation than TKAs revised for other reasons.

Methods: A total of 73 explanted fixed-bearing TKAs (42 CR and 31 PS) were examined.

View Article and Find Full Text PDF

Studies have demonstrated that at least 20% of individuals infected with SARS-CoV-2 remain asymptomatic. Although most global efforts have focused on severe illness in COVID-19, examining asymptomatic infection provides a unique opportunity to consider early immunological features that promote rapid viral clearance. Here, postulating that variation in the human leukocyte antigen (HLA) loci may underly processes mediating asymptomatic infection, we enrolled 29,947 individuals, for whom high-resolution HLA genotyping data were available, in a smartphone-based study designed to track COVID-19 symptoms and outcomes.

View Article and Find Full Text PDF

As the number of young and active individuals undergoing Total Hip Arthroplasty (THA) are increasing yearly, there is a need for hip prostheses to have increased longevity. Current investigations into the longevity of these prostheses only include walking as the patient's activity as there is limited data on the amount and intensity of other activity performed by the patient. To further understand the evolution of wear and increase the longevity of these implants, the impact of different activities on the hip prosthesis needs to be investigated.

View Article and Find Full Text PDF

Hip joints with bearings composed of cobalt-chromium alloy (metal-on-metal bearings) have been one of the most widely used implants in joint replacement arthroplasty. Unfortunately, these implants can contribute to a complication called aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL), a type IV metal hypersensitivity response around the joint. Consistent with such bearings, increased metal debris can be found in the surrounding fluids and in remote tissues and organs, due to wear and corrosion.

View Article and Find Full Text PDF

Background: Over five million joint replacements are performed across the world each year. Cobalt chrome (CoCr) components are used in most of these procedures. Some patients develop delayed-type hypersensitivity (DTH) responses to CoCr implants, resulting in tissue damage and revision surgery.

View Article and Find Full Text PDF

Aims: The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs).

Methods: At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface.

View Article and Find Full Text PDF

Despite some inconsistent reporting of symptoms, studies have demonstrated that at least 20% of individuals infected with severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) will remain asymptomatic. Although most global efforts have focused on understanding factors underlying severe illness in COVID-19 (coronavirus disease of 2019), the examination of asymptomatic infection provides a unique opportunity to consider early disease and immunologic features promoting rapid viral clearance. Owing to its critical role in the immune response, we postulated that variation in the human leukocyte antigen (HLA) loci may underly processes mediating asymptomatic infection.

View Article and Find Full Text PDF

The impact of COVID-19 varies markedly, not only between individual patients but also between different populations. We hypothesised that differences in human leukocyte antigen (HLA) genes might influence this variation. Using next generation sequencing, we analysed the class I and class II classical HLA genes of 147 individuals of European descent experiencing variable clinical outcomes following COVID-19 infection.

View Article and Find Full Text PDF

The generation of particulate debris at the taper junction of total hip replacements (THRs), can cause failure of the artificial hip. The taper surfaces of femoral heads and trunnions of femoral stems are generally machined to a certain roughness to enhance fixation. However, the effect of the surface roughness of these surfaces on the fixation, wear and consequently clinical outcomes of the design is largely unknown.

View Article and Find Full Text PDF

Total hip replacement (THR) is one of the most successful orthopaedic surgeries; however, failures can occur due to adverse reactions to wear debris. Recently, a large number of failures linked to the release of metal particles from the taper junction between femoral head and femoral stem have been reported. One possible reason for this may be design variations such as taper mismatches associated with the taper and trunnion angles.

View Article and Find Full Text PDF

Objectives: To determine risk factors for revision in patients implanted with a commonly used metal on metal (MoM) hip replacement.

Design: Retrospective cohort study in combination with a prospective national retrieval study (Northern Retrieval Registry (NRR)).

Setting: Combined orthopaedic unit in combination with the NRR.

View Article and Find Full Text PDF

A recognized issue related to metal-on-metal hip resurfacings is early fracture of the femur. Most theories regarding the cause of fracture relate to clinical factors but an engineering analysis of failed hip resurfacings has not previously been reported. The objective of this work was to determine the wear volumes and surface roughness values of a cohort of retrieved hip resurfacings which were removed due to early femoral fracture, infection and avascular necrosis (AVN).

View Article and Find Full Text PDF

Taper wear at the head-neck junction is a possible cause of early failure in large head metal-on-metal (LH-MoM) hip replacements. We hypothesized that: (i) taper wear may be more pronounced in certain product designs; and (ii) an increased abductor moment arm may be protective. The tapers of 104 explanted LH-MoM hip replacements revised for adverse reaction to metal debris (ARMD) from a single manufacturer were analyzed for linear and volumetric wear using a co-ordinate measuring machine.

View Article and Find Full Text PDF

Hip replacements are used to improve the quality of life of people with orthopaedic conditions, but the use of metal-on-metal (MoM) arthroplasty has led to poor outcomes for some patients. These problems are related to the generation of micro- to nanosized metal wear particles containing Cr, Co or other elements, but the current analytical methods used to investigate the processes involved do not provide sufficient information to understand the size or composition of the wear particles generated in vivo. In this qualitative feasibility study, asymmetric flow field-flow fractionation (AF(4)) coupled with inductively coupled plasma mass spectrometry (ICP-MS) was used to investigate metal protein binding and the size and composition of wear metal particles present in serum and hip aspirates from MoM hip replacement patients.

View Article and Find Full Text PDF

In contrast to the articulating and taper surfaces of failed total hip replacements, volumetric wear analysis of trunnions is not routinely performed. Metal wear particles from the trunnion may contribute not only to the failures of metal-on-metal total hip replacements but also to all hip replacements utilising metal trunnions. A validation study was performed with the material removed in stages from the trunnions of an Exeter V40 stem, a Corail stem and an Accolade stem to simulate different magnitudes of wear.

View Article and Find Full Text PDF

Objective: To determine whether elevated blood cobalt (Co) concentrations are associated with early failure of metal-on-metal (MoM) hip resurfacings secondary to adverse reaction to metal debris (ARMD).

Design: Cohort study.

Setting: Single centre orthopaedic unit.

View Article and Find Full Text PDF

Aim: To describe the histopathology of localised adverse reactions to metal debris (ARMD) seen in association with failed metal on metal (MoM) hip arthroplasties. The nature of aseptic lymphocytic vasculitis associated lesion (ALVAL) is investigated.

Methods: Periprosthetic soft tissues biopsied at time of revision from failed MoM hip arthroplasties from January 2007 to March 2011 were analysed.

View Article and Find Full Text PDF