Publications by authors named "David Cembrero-Fucinos"

The verification of examination procedures is a responsibility for clinical laboratories in order to guarantee that their performance characteristics comply with the specifications obtained during the validation process and are congruent with the intended scope of the assay. The aim was to perform an evaluation of precision, bias, linearity, linear drift, sample carry-over, and comparability of 73 assays from Siemens Healthineers, by following the CLSI EP10-A3 guidelines. The verification was performed by measuring 72 biochemical parameters in quality control (QC) materials from Bio-Rad (except for IL6) with 73 assays installed on eight measuring systems (five Atellica CH 930 and three IM 1600 analyzers from Siemens Healthcare Diagnostics).

View Article and Find Full Text PDF

Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) has been suggested as a reliable method for bacterial identification from cultures. Direct analysis of clinical samples might increase the usefulness of this method, shortening the time for microorganism identification. We compared conventional methods for the diagnosis of urinary tract infections (UTIs) and identification of the urinary tract pathogens (automated screening, plate cultures, and identification based on biochemical characteristics) and a fast method based on conventional screening and MALDI-TOF MS.

View Article and Find Full Text PDF