Publications by authors named "DM Lovelock"

Article Synopsis
  • * A study evaluated the myQA SRS device, a high-resolution CMOS 2D detector, for its performance in PSQA by characterizing its dosimetric properties and conducting gamma index analyses on nine simulated and clinical plans.
  • * Results indicated that the myQA SRS showed excellent dosimetric agreement with dosimetry films, achieving average gamma index pass rates of 99.9% and 99.2%, demonstrating its reliability and potential as a modern alternative to traditional film dosim
View Article and Find Full Text PDF

Purpose: In radiation therapy (RT), if an immobilization device is lost or damaged, the patient may need to be brought back for resimulation, device fabrication, and treatment planning, causing additional imaging radiation exposure, inconvenience, cost, and delay. We describe a simulation-free method for replacing lost or damaged RT immobilization devices.

Methods And Materials: Replacement immobilization devices were fabricated using existing simulation scans as design templates by computer numerical control (CNC) milling of molds made from extruded polystyrene (XPS).

View Article and Find Full Text PDF

Background: Electromagnetic transponders bronchoscopically implanted near the tumor can be used to monitor deep inspiration breath hold (DIBH) for thoracic radiation therapy (RT). The feasibility and safety of this approach require further study.

Methods: We enrolled patients with primary lung cancer or lung metastases.

View Article and Find Full Text PDF

Purpose: To evaluate the efficacy of using bronchoscopically implanted anchored electromagnetic transponders (EMTs) as surrogates for 1) tumor position and 2) repeatability of lung inflation during deep-inspiration breath-hold (DIBH) lung radiotherapy.

Methods: Forty-one patients treated with either hypofractionated (HF) or conventional (CF) lung radiotherapy on an IRB-approved prospective protocol using coached DIBH were evaluated for this study. Three anchored EMTs were bronchoscopically implanted into small airways near or within the tumor.

View Article and Find Full Text PDF

Introduction: Paraspinal stereotactic body radiotherapy (SBRT) involves risks of severe complications. We evaluated the safety of the paraspinal SBRT program in a large academic hospital by applying failure modes and effects analysis.

Methods: The analysis was conducted by a multidisciplinary committee (two therapists, one dosimetrist, four physicists, and two radiation oncologists).

View Article and Find Full Text PDF

Deep inspiration breath hold (DIBH) has dosimetric advantages for lung cancer patients treated with external beam therapy, but is difficult for many patients to perform. Proton therapy permits sparing of the downstream organs at risk (OAR). We compared conventionally fractionated proton (p) and photon(x) plans on both free breathing (FB) and DIBH planning CTs to determine the effect of DIBH with proton therapy.

View Article and Find Full Text PDF

Managing TG-51 reference dosimetry in a large hospital network can be a challenging task. The objectives of this study are to investigate the effectiveness of using Statistical Process Control (SPC) to manage TG-51 workflow in such a network. All the sites in the network performed the annual reference dosimetry in water according to TG-51.

View Article and Find Full Text PDF

Objective: Adaptive radiation therapy (ART) is an integral part of MR-guided RT (MRgRT), requiring a new RT plan for each treatment fraction and resulting in a significant increase in patient-specific quality assurance (PSQA). This study investigates the possibility of using treatment log-file for automated PSQA.

Method: All treatment plans were delivered in 1.

View Article and Find Full Text PDF

Purpose/objective(s): The additional personnel and imaging procedures required for Adaptive Radiation Therapy (ART) pose a challenge for a broad implementation. We hypothesize that a change in transit fluence during the treatment course is correlated with the change of quality of life and thus can be used as a replanning trigger.

Materials/methods: Twenty-one head and neck cancer (HNC) patients filled out an MD Anderson Dysphagia Inventory (MDADI) questionnaire, before-and-after the radiotherapy treatment course.

View Article and Find Full Text PDF

Purpose: This study aimed to evaluate and compare different system calibration methods from a large cohort of systems to establish a commissioning procedure for surface-guided frameless cranial stereotactic radiosurgery (SRS) with intrafractional motion monitoring and gating. Using optical surface imaging (OSI) to guide non-coplanar SRS treatments, the determination of OSI couch-angle dependency, baseline drift, and gated-delivered-dose equivalency are essential.

Methods: Eleven trained physicists evaluated 17 OSI systems at nine clinical centers within our institution.

View Article and Find Full Text PDF

The use of trajectory log files for routine patient quality assurance is gaining acceptance. Such use requires the validation of the trajectory log itself. However, the accurate localization of a multileaf collimator (MLC) leaf while it is in motion remains a challenging task.

View Article and Find Full Text PDF

Purpose: We sought to investigate the tumor control probability (TCP) of spinal metastases treated with stereotactic body radiation therapy (SBRT) in 1 to 5 fractions.

Methods And Materials: PubMed-indexed articles from 1995 to 2018 were eligible for data extraction if they contained SBRT dosimetric details correlated with actuarial 2-year local tumor control rates. Logistic dose-response models of collected data were compared in terms of physical dose and 3-fraction equivalent dose.

View Article and Find Full Text PDF

In adaptive radiation therapy of head and neck cancer, any significant anatomical changes observed are used to adapt the treatment plan to maintain target coverage without elevating the risk of xerostomia. However, the additional resources required for adaptive radiation therapy pose a challenge for broad-based implementation. It is hypothesized that a change in transit fluence is associated with volumetric change in the vicinity of the target and therefore can be used as a decision support metric for adaptive radiation therapy.

View Article and Find Full Text PDF

Spine stereotactic body radiation therapy frequently involves the delivery of high doses to targets in proximity to the spinal cord; thus, the radiation must be delivered with great spatial accuracy. Monitoring for large shifts in target and cord position that might occur during dose delivery is a challenge for clinics equipped with a conventional C-arm Linac. Treatment must be halted, then imaging and registration must be done to determine whether a significant shift has occurred.

View Article and Find Full Text PDF

Introduction: Often in clinical practice radiation oncologists encounter patients who require treatment to the spine commonly in the setting of metastatic disease. These metastases usually cause pain, immobility, or neurologic deficits mandating expedited therapy to alleviate the suffering of our patients. Spine radiosurgery techniques have been used extensively for palliation purposes; however, given the patients' deteriorating condition or pain and inability to tolerate anesthesia the radiation oncologist is often left with the conundrum of how to best set up his or her patient in preparation for radiosurgery if supine is not a viable option.

View Article and Find Full Text PDF

OBJECTIVE The aim of this study was to evaluate the safety and efficacy of kyphoplasty treatment prior to spine stereotactic radiosurgery (SRS) in patients with spine metastases. METHODS A retrospective review of charts, radiology reports, and images was performed for all patients who received SRS (single fraction; either standalone or post-kyphoplasty) at a large tertiary cancer center between January 2012 and July 2015. Patient and tumor variables were documented, as well as treatment planning data and dosimetry.

View Article and Find Full Text PDF

Purpose: An ideal commissioning and quality assurance (QA) program for Volumetric Modulated Arc Therapy (VMAT) delivery systems should assess the performance of each individual dynamic component as a function of gantry angle. Procedures within such a program should also be time-efficient, independent of the delivery system and be sensitive to all types of errors. The purpose of this work is to develop a system for automated time-resolved commissioning and QA of VMAT control systems which meets these criteria.

View Article and Find Full Text PDF

Purpose: To investigate constancy, within a treatment session, of the time lag relationship between implanted markers in abdominal tumors and an external motion surrogate.

Methods: Six gastroesophageal junction and three pancreatic cancer patients (IRB-approved protocol) received two cone-beam CTs (CBCT), one before and one after treatment. Time between scans was less than 30 min.

View Article and Find Full Text PDF

Purpose: To compare the potential benefits of continuous monitoring of prostate position and intervention (CMI) using 2-mm displacement thresholds during stereotactic body radiation therapy (SBRT) treatment to those of a conventional image-guided procedure involving single localization prior to treatment.

Methods And Materials: Eighty-nine patients accrued to a prostate SBRT dose escalation protocol were implanted with radiofrequency transponder beacons. The planning target volume (PTV) margin was 5 mm in all directions, except for 3 mm in the posterior direction.

View Article and Find Full Text PDF

Purpose: Certain types of commonly used fiducial markers take on irregular shapes upon implantation in soft tissue. This poses a challenge for methods that assume a predefined shape of markers when automatically tracking such markers in kilovoltage (kV) radiographs. The authors have developed a method of automatically tracking regularly and irregularly shaped markers using kV projection images and assessed its potential for detecting intrafractional target motion during rotational treatment.

View Article and Find Full Text PDF

Object: Radiation therapy is known to impair wound healing. Higher dose per fraction is believed to increase this risk. This study sought to quantify rates of wound complication in patients receiving preoperative conventionally fractionated radiotherapy (XRT) or high-dose hypofractionated image-guided radiation therapy (IGRT) for spinal metastasis, and to identify predictors of wound complication.

View Article and Find Full Text PDF

Purpose: Abdominal compression using a pneumatic abdominal compression belt developed in-house has been used to reduce respiratory motion of patients undergoing hypo-fractionated or single fraction stereotactic radio-ablative therapy for abdominal cancers. The clinical objective of belt usage was to reduce the cranial-caudal (CC) respiratory motion of the tumor to 5 mm or less during both CT simulation and treatment. A retrospective analysis was done to determine the effectiveness of the device and associated clinical procedures to reduce the CC respiratory motion of the tumor.

View Article and Find Full Text PDF

To provide an alternative device for immobilization of the head while easing claustrophobia and improving comfort, an "open-face" thermoplastic mask was evaluated using video-based optical surface imaging (OSI) and kilovoltage (kV) X-ray radiography. A three-point thermoplastic head mask with a precut opening and reinforced strips was developed. After molding, it provided sufficient visible facial area as the region of interest for OSI.

View Article and Find Full Text PDF

Background: En bloc wide-margin excision significantly decreases the risk of chordoma recurrence. However, a wide surgical margin cannot be obtained in many chordomas because they arise primarily in the sacrum, clivus, and mobile spine. Furthermore, these tumors have shown resistance to fractionated photon radiation at conventional doses and numerous chemotherapies.

View Article and Find Full Text PDF

Purpose: Commercial CT-based image-guided radiotherapy (IGRT) systems allow widespread management of geometric variations in patient setup and internal organ motion. This document provides consensus recommendations for quality assurance protocols that ensure patient safety and patient treatment fidelity for such systems.

Methods: The AAPM TG-179 reviews clinical implementation and quality assurance aspects for commercially available CT-based IGRT, each with their unique capabilities and underlying physics.

View Article and Find Full Text PDF