Publications by authors named "Corey Clift"

RadBall™ is a novel technology that can locate unknown radioactive hazards within contaminated areas, hot cells, and gloveboxes. The device consists of a colander-like outer tungsten collimator that houses a radiation-sensitive polymer semisphere. The collimator has a number of small holes; as a result, specific areas of the polymer are exposed to radiation, becoming increasingly more opaque in proportion to the absorbed dose.

View Article and Find Full Text PDF

This work presents an investigation into the use of PRESAGE dosimeters with an optical-CT scanner as a 3D dosimetry system for quantitative verification of respiratory-gated treatments. The CIRS dynamic thorax phantom was modified to incorporate a moving PRESAGE dosimeter-simulating respiration motion in the lungs. A simple AP/PA lung treatment plan was delivered three times to the phantom containing a different but geometrically identical PRESAGE insert each time.

View Article and Find Full Text PDF

Achieving accurate small field dosimetry is challenging. This study investigates the utility of a radiochromic plastic PRESAGE read with optical-CT for the acquisition of radiosurgery field commissioning data from a Novalis Tx system with a high-definition multileaf collimator (HDMLC). Total scatter factors (Sc, p), beam profiles, and penumbrae were measured for five different radiosurgery fields (5, 10, 20, 30 and 40 mm) using a commercially available optical-CT scanner (OCTOPUS, MGS Research).

View Article and Find Full Text PDF