Despite measures to reduce disease transmission, a risk can occur when blood glucose meters (BGMs) are used on multiple individuals or by caregivers assisting a patient. The laboratory and in-clinic performance of a BGM system before and after disinfection should be demonstrated to guarantee accurate readings and reliable control of blood glucose (BG) for patients. In this study, an effective disinfection procedure, conducting wiping 10 times to assure a one minute contact time of the disinfectant on contaminated surface, was first demonstrated using test samples of the meter housing materials, including acrylonitrile butadiene styrene (ABS), polymethyl methacrylate (PMMA), and polycarbonate (PC), in accordance with ISO 15197:2013.
View Article and Find Full Text PDF