Background: New oscillometry devices allowing quantification of respiratory function using tidal breathing are commercially available, but reference equations are lacking for the multiethnic Canadian pediatric population.
Methods: We conducted a prospective cross-sectional study of healthy children carefully selected for absence of asthma, atopy, tobacco smoke, obesity, prematurity, and recent respiratory infection. Triplicate measures were obtained of respiratory system resistance (Rrs) and reactance (Xrs), area under the reactance curve (AX) and resonant frequency (Fres) on four signals, whose testing order was randomized: two signals on the Resmon Pro Full (8 Hz and 5-11-19 Hz) and two signals on the tremoflo C-100 (5-37 Hz and 7-41 Hz).
Background: With several commercially available devices measuring respiratory impedance by oscillometry, the agreement between values obtained on different instruments or frequencies remains unclear. Our aim was to examine the agreement between resistance and reactance parameters on two oscillometry instruments using different waveforms.
Methods: We conducted a prospective cross-sectional study in asthmatic children aged 3-17 years.