Background: Metal-on-metal total hip prostheses will produce a certain amount of wear debris. This results in increased whole-blood metal levels, which may cause adverse effects. It is not known to what extent the problem has been overcome by advances in alloy technology.

View Article and Find Full Text PDF

A tribologic assessment was performed on 22 metal-metal hip prostheses from a single manufacturer, following removal for early aseptic loosening after a mean service life of 32 months (range, 12-59 months). The mean linear wear rate was 7.6 microm/year (range, 2.