In light of the proposal from the International Commission on Radiological Protection for a lowered eye dose limit, now adopted by a European Union Council Directive, lead glasses may be required for some staff in interventional radiology to ensure that occupational exposure is as low as reasonably practicable. To investigate the lens protection offered from various models of lead glasses exposed to X-rays coming from a source to the left and below, calibrated radiochromic film was positioned in the lens area of a head phantom. When the source-to-eye angles were large, the dose reduction factors (the ratio of eye dose without protection to dose with protection) to the right lens area were much lower than to the left lens area, particularly with smaller-lensed glasses, due to gaps in protection between the face and the glasses.
View Article and Find Full Text PDFObjectives: To assess the cross-centre consistency of iodine enhancement, contrast-to-noise ratio and radiation dose in a multicentre perfusion CT trial of colorectal cancer.
Materials And Methods: A cylindrical water phantom containing different iodine inserts was examined on seven CT models in 13 hospitals. The relationship between CT number (Hounsfield units, HU) and iodine concentration (milligrams per millilitre) was established and contrast-to-noise ratios (CNRs) calculated.