Because a case report and theoretical mass balances suggested that hypoxic guard systems may not prevent the formation of hypoxic inspired mixtures (FIO2 ≤ 21 %) over the clinically used fresh gas flow (FGF) range, we measured FIO2 over a wide range of hypoxic guard limits for O2/N2O and O2/air mixtures. After IRB approval, 16 ASA I-II patients received sevoflurane in either O2/N2O (n = 8) or O2/air (n = 8) using a Zeus(®) anesthesia machine in the conventional mode. After using an 8 L/min FGF with FDO2 = 25% for 10 min, the following hypoxic guard limits were tested for 4 min each, expressed as [total FGF in L/min; FDO2 in %]: [0.