Purpose: The aim of this study is the experimental and Monte Carlo-based determination of small field correction factors for the unshielded silicon detector microSilicon for a standard linear accelerator as well as the Cyberknife System. In addition, a detailed Monte Carlo analysis has been performed by modifying the detector models stepwise to study the influences of the detector's components.
Methods: Small field output correction factors have been determined for the new unshielded silicon diode detector, microSilicon (type 60023, PTW Freiburg, Germany) as well as for the predecessors Diode E (type 60017, PTW Freiburg, Germany) and Diode SRS (type 60018, PTW Freiburg, Germany) for a Varian TrueBeam linear accelerator at 6 MV and a Cyberknife system.
In the present study, surface-modified nanoparticles based on biodegradable material were used for antibody coupling in order to get a selective drug carrier systems. Gelatin nanoparticles were prepared by a desolvation process. Sulfhydryl groups were introduced which enabled the linkage of NeutrAvidin (NAv).
View Article and Find Full Text PDF