Purpose: Small animal irradiators are equipped with x-ray beams and cone collimators with millimeter dimensions to be used in preclinical research. The use of small fields in the kV energy range may require the application of energy-dependent, field size-dependent, or depth-dependent correction factors to the dosimetric data acquired for treatment planning system (TPS) commissioning purposes to obtain accurate dose values. Considering that these corrections are also detector dependent, the suitability of a synthetic single-crystal diamond detector for small-field relative dosimetry in a preclinical irradiator (220-kVp) was evaluated to avoid the necessity of applying correction factors during TPS commissioning.