Clin Oral Investig
November 2024
Objectives: To investigate the influence of combining levels of metal artifact reduction (MAR) tool with different filters and slice-thickness in the detection of mandibular canals perforated by implants on CBCT scans and to objectively assess the impact of the aforementioned combinations.
Materials And Methods: Implants were placed above (8 implants) and inside (10 implants) mandibular canals of dried-mandibles. CBCT scans were obtained with the Eagle 3D unit (85 kVp, 8 mA, 5 × 5 cm FOV, 130 μm voxel size, and off/medium/high MAR levels).
Objectives: To evaluate the trueness of dental arches digitised by two intraoral scanning (IOS) technologies from patients presenting crowding, diastema, and bilateral posterior edentulous space with tilted molar.
Methods: Conventional impressions and dental stone models were generated from three patients presenting the aforementioned dental arch conditions. These models were digitised on a desktop scanner, and the resulting mesh was used as reference.
Objectives: This study evaluated the effect of metal artefact reduction (MAR) level and tube current on the assessment of dental implant positioning relative to the mandibular canal (MC) through cone-beam computed tomography (CBCT).
Methods: Titanium dental implants were placed in dried mandibles at 0.5-mm superior to the MC (group 1/n = 8) and 0.