Publications by authors named "Caitlin DiGeorgio"

Background: Little guidance exists regarding the minimum screw length and screw quantity necessary to achieve fixation in reverse total shoulder arthroplasty (rTSA); to that end, this study quantified the displacement of 2 different sizes of glenoid baseplates using multiple different screw lengths and quantities of screws in a low-density polyurethane bone-substitute model.

Methods: Testing of rTSA glenoid loosening was conducted according to ASTM F 2028-17. To independently evaluate the impact of screw quantity and screw length on rTSA glenoid fixation for 2 different sizes of glenoid baseplates, baseplates were constructed using 2 screws, 4 screws, or 6 screws (with the latter being used for the larger baseplate only) with 3 different poly-axial locking compression screw lengths.

View Article and Find Full Text PDF