Medicinal aerosol products (metered dose and dry powder inhalers) require characterization testing over a wide range of use and pre-operating stress scenarios in order to ensure robust product performance and support submissions for regulatory approval. Aerosol characterization experiments on Staccato(®) Loxapine for inhalation (Staccato Loxapine) product (emitted dose, particle size, and purity) were assessed at different operating settings (flow rates, ambient temperature and humidity, altitude, and orientation) and at nominal test conditions following exposure to various stresses on the device (mechanical shock, vibration, drop, thermal cycling, and light exposure). Emitted dose values were approximately 90% of the coated dose at every condition, meeting target specifications in each case.