The performance of the SE-9000 automated haematology analyser in a laboratory receiving a high number of abnormal specimens from haematological oncology patients was assessed according to formal protocols for the evaluation of blood cell counters. Linearity over a useful working range, precision in clinically important ranges and negligible carry-over were demonstrated in this group of patient samples confirming the results of previous investigators. The comparability of instrument derived differential leucocyte counts from both normal and distributionally abnormal samples with those obtained by visual microscopy using the NCCLS H-20 A protocol was very good.