In this study, an accurate analytical method development for cobalt determination in urine samples was described. The method is based on the mass transfer of the target analytes to the organic phase from the aqueous phase by the dispersing extractant throughout the solution with the aid of CO bubbles prior to sample measurement by using a slotted quartz tube flame atomic absorption spectrophotometer. An extractor (1-decanol) dropped effervescent tablet (anhydrous sodium carbonate and sodium dihydrogen phosphate dihydrate mixture) was used in order to separate/preconcentrate cobalt after complexation of cobalt ions in aqueous solution with the Schiff base ligand.