A rapid calibration system is under development for the Near Backscatter Imager (NBI) in use at the National Ignition Facility (NIF). NBI is an optical diagnostic that quantifies the stimulated Brillouin and Raman backscatter produced by NIF's targets. Specifically, NBI measures the light that does not fall directly back into the laser aperture, which is measured by the Full Aperture Backscatter System (FABS).
View Article and Find Full Text PDFPhys Rev Lett
September 2022
Measurements were made of the return current instability growth rate, demonstrating its concurrence with nonlocal transport. Thomson scattering was used to measure a maximum growth rate of 5.1×10^{9} Hz, which was 3 times less than classical Spitzer-Härm theory predicts.
View Article and Find Full Text PDFThomson scattering measurements in high energy density experiments are often recorded using optical streak cameras. In the low-signal regime, noise introduced by the streak camera can become an important and sometimes the dominant source of measurement uncertainty. In this paper, we present a formal method of accounting for the presence of streak camera noise in our measurements.
View Article and Find Full Text PDFOptical collective Thomson scattering (TS) is used to diagnose magnetized high energy density physics experiments at the Magpie pulsed-power generator at Imperial College London. The system uses an amplified pulse from the second harmonic of a Nd:YAG laser (3 J, 8 ns, 532 nm) to probe a wide diversity of high-temperature plasma objects, with densities in the range of 10-10 cm and temperatures between 10 eV and a few keV. The scattered light is collected from 100 μm-scale volumes within the plasmas, which are imaged onto optical fiber arrays.
View Article and Find Full Text PDFWe present the first local, quantitative measurements of ion current filamentation and magnetic field amplification in interpenetrating plasmas, characterizing the dynamics of the ion Weibel instability. The interaction of a pair of laser-generated, counterpropagating, collisionless, supersonic plasma flows is probed using optical Thomson scattering (TS). Analysis of the TS ion-feature revealed anticorrelated modulations in the density of the two ion streams at the spatial scale of the ion skin depth c/ω_{pi}=120 μm, and a correlated modulation in the plasma current.
View Article and Find Full Text PDF