This study examined the feasibility of applying frequency-modulated spectroscopy (FMS) to test vacuum seal integrity of lyophilized protein pharmaceuticals in glass vials. A lyophilized recombinant monoclonal antibody was used as an example to demonstrate that FMS is a non-destructive method that could accurately and quickly determine vial vacuum integrity within a pressure range of 0.04 to 0.