Int Forum Allergy Rhinol
September 2019
Background: Quantitative olfactory testing is essential to determine the validity and nature of a patient's complaint, accurately monitor changes in function over time (including influences of pharmacological, surgical, or immunological interventions), detect malingering, and establish disability compensation. This work describes the clinical validation of an advanced rapid odor detection threshold test that standardizes odorant delivery and eliminates unsanitary sniff bottles, the need for blindfolds, and the possibility of touching an odorant stimulus to the nose.
Methods: Snap & Sniff® single-staircase phenyl ethyl alcohol detection thresholds (S&S-Ts) were assessed bilaterally in 736 patients with chemosensory complaints; 421 received both bilateral and unilateral testing.