Introduction: The aim of this study was to perform a verification of the Dymind D7-CRP automated analyser and compare it with established analysers.
Materials And Methods: Analytical verification included estimation of repeatability, between run precision, within-laboratory precision, and bias in control samples with low, normal and high levels. The acceptance criteria for analytical verification were defined using the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) 2019 Biological Variation Database.
Background: Results from hemolyzed, icteric, and lipemic samples may be inaccurate and can lead to medical errors. These preanalytical interferences may be detected using visual or automated assessment. Visual inspection is time consuming, highly subjective and not standardized.